TSI AEROTRAK

TSI · AEROTRAK · 72xx,73xx,75xx

TSI AEROTRAK
Manufacturer
TSI
Product
AEROTRAK
Model
72xx,73xx,75xx
Protocol
Modbus
Updated
2026-05-28
Driver ID
142

Notes

Particle Counter for clean room applications.

Driver supports Modbus/TCP only.

Default IP address of Aerotrak is 192.168.200.90.

To set up Aerotrak, use TSI configuration software. 
Go to Tech Page, and make sure the Modbus Map Version is set to v15.
Also set time and IP address.

Registers 15

#NameUnitDescription
1Status
2Sample countcount
3DateYMD
4TimeHMS
5Channel 1count
6Channel 2count
7Channel 3count
8Channel 4count
9Location ID
10Sample times
11Hold times
12Flow rateL/min
13Relative humidity%
14Temperature°C
15Status